Monte Carlo calculated CT numbers for improved heavy ion treatment planning
Better knowledge of CT number values and their uncertainties can be applied to improve heavy ion treatment
planning. We developed a novel method to calculate CT numbers for a computed tomography (CT) scanner
using the Monte Carlo (MC) code, BEAMnrc/EGSnrc. To generate the initial beam shape and spectra we conducted
full simulations of an X-ray tube, filters and beam shapers for a Siemens Emotion CT. The simulation output files were
analyzed to calculate projections of a phantom with inserts. A simple reconstruction algorithm (FBP using a Ram-Lak
filter) was applied to calculate the pixel values, which represent an attenuation coefficient, normalized in such a way
to give zero for water (Hounsfield unit (HU)). Measured and Monte Carlo calculated CT numbers were compared.
The average deviation between measured and simulated CT numbers was 4 ± 4 HU and the standard deviation σ was
49 ± 4 HU. The simulation also correctly predicted the behaviour of H-materials compared to a Gammex tissue substitutes.
We believe the developed approach represents a useful new tool for evaluating the effect of CT scanner and
phantom parameters on CT number values.